American BioTech Supply 72 Cu.Ft. Premier Laboratory Refrigerator, 1°C - 10°C, Swing Glass Doors, ABT-HC-LP-72
The Premier Laboratory Refrigerators feature a microprocessor temperature controller and display, allowing for precise temperature control, verification and recovery versus mechanical style thermostats. An interior fan driven, forced-draft air circulation system provides excellent temperature uniformity and recovery after door openings. Critical samples and supplies are security backed with an extensive package of audible and visual alarms.
The temperature display and alarm module offers an array of robust security features with a monitoring and alarm system that is 100% independent of a microprocessor temperature controller.
Environmentally friendly, natural hydrocarbon (HC) refrigerants (not containing hydrofluorocarbons) vastly reduce global-warming potential (GWP) while meeting new EPA/SNAP mandates and UL, ASHRAE compliance guidelines.
Details:
- 72 CU. FT. CAPACITY
- THREE SWING GLASS DOORS, SELF-CLOSING, RIGHT HINGED, NOT REVERSIBLE
- TWELVE ADJUSTABLE SHELVES (ADJUSTABLE IN ½" INCREMENTS)
- FORCED DRAFT AIR CIRCULATION
- KEYED DOOR LOCK
- MAGNETIC DOOR GASKET FOR POSITIVE SEAL
- HEAVY DUTY DOOR HANDLE
- LED INTERIOR LIGHTS ARE SAFETY SHIELDED AND SWITCH CONTROLLED
- SWIVEL CASTERS (3 ½”)
- PROBE ACCESS PORT (3/8")
- WHITE POWDER-COATED STEEL EXTERIOR AND INTERIOR
- HIGH DENSITY URETHANE FOAM CABINET INSULATION
- HYDROCARBON, NATURAL REFRIGERANT (R290)
- ETL LISTED
- TWO YEAR PARTS AND LABOR WARRANTY, PLUS AN ADDITIONAL THREE YEAR COMPRESSOR PARTS WARRANTY
Specifications:
- EXTERIOR DIMENSIONS: 81” W X 34” D X 78 ¼” H WITHOUT CASTERS; 81 ¾” H WITH CASTERS
- CYCLE DEFROST
- 115V, 60 HZ, 7.5 AMPS, 1/2 HP
- SHIPPING WEIGHT: 743 LBS.
Temp Controller:
- MICROPROCESSOR TEMPERATURE CONTROLLER
- ADJUSTABLE OPERATING TEMPERATURE RANGE: 1°-10°C
- DIGITAL TEMPERATURE DISPLAY AND ALARM MODULE
- SAMPLE SIMULATED DIGITAL TEMPERATURE DISPLAY; .1° RESOLUTION ADJUSTABLE TEMPERATURE DISPLAY IN CELSIUS OR FAHRENHEIT
- AUDIBLE AND VISUAL HIGH/LOW TEMPERATURE ALARMS
- 2 PROBES; 1 IN AIR AND 1 IN SAMPLE BOTTLE
- SENSOR ERROR ALARM
- POWER FAILURE ALARM
- RESETTABLE 1-POINT MIN/MAX HISTORY
- REMOTE ALARM CONTACTS
- ALARM VALIDATION (AUDIBLE, VISUAL, AND REMOTE ALARM)
- ALARM MUTE WITH RING BACK